Products Description
Emerson Rosemount X-STREAM Gas Analyzer Unit

Module Feature
X-STREAM is the modular continuous gas analysis platform from Emerson Rosemount Analytical.
It supports up to five gas components measured simultaneously, with multiple optional measurement principles including NDIR infrared, UV/visible photometry, paramagnetic oxygen, thermal conductivity and electrochemical sensors. It adopts IntrinzX optical technology to improve long-term stability and reduce calibration frequency. It has multiple housing variants: XEGP rack general-purpose type, XEXF field wall-mount type and XEFD flameproof hazardous-area type, widely used for boiler combustion monitoring, petrochemical process, flue gas emission monitoring and gas plant analysis cabinets.
Working principle
The built-in optical cell and sensing unit absorb specific wavelength light corresponding to target gas.
The microprocessor calculates gas concentration based on light attenuation or paramagnetic characteristics. It records measurement data, calibration logs and fault events, supports web page access, Modbus and digital signal upload to upper DCS/SCADA control system.
Onboard components
Optical photometry module, sensor sampling cell, main processing board, local LCD HMI, sample flow control unit, signal isolation circuit, power supply module.
LED indicators
Front LEDs display power status, measurement running state, calibration progress and fault alarm.
Hardware self‑diagnosis
Continuous self-check detects sample flow loss, optical contamination, sensor failure, internal temperature anomaly and communication interruption, alarm messages display on local screen and web interface.
Programming software
Emerson Rosemount Analytical Configuration Tool, used for gas channel setup, calibration, alarm threshold configuration and data log export.
Electrical & Process Parameters
Model:X-STREAM Brand:Emerson Rosemount Analytical Device type:Multi-component continuous gas analyzer Supply voltage:100–240V AC Function:Process gas, flue gas concentration continuous measurement, combustion optimization and emission monitoring
Mechanical & Structure
Mounting:19-inch rack or wall flange mounting
Protection class:IP20 / IP66 optional
Housing:Aluminum / stainless steel analyzer enclosure
Environmental
Operating temperature:-20℃ ~ +50℃ Storage temperature:-40℃ ~ +85℃ Ambient humidity:5~95% non‑condensing
Certification:CE, UL, CSA
Status:Original Emerson process analysis spare
Installation
Install analyzer inside cabinet or field wall bracket, connect sample gas pipeline, drain and calibration gas circuit.
Set gas component parameters and perform zero/span calibration, verify sample flow and output signal before commissioning.
Interface
Sample gas inlet/outlet, 4-20mA analog terminals, digital I/O, Ethernet communication port.
Diagnostic
Measurement drift: Dirty optical cell, contaminated sample filter or expired calibration gas.
Flow fault: Blocked sample line, damaged sampling pump.
Communication offline: Disconnected network cable or wrong IP address.

Product Brief
Emerson Rosemount X-STREAM multi-component gas analyzer, modular on-line instrument for boiler flue gas and petrochemical process gas continuous monitoring.